TR Tester, ET 01

TR Tester, ET 01-II, for determination of low temperature characteristics by the temperature retraction procedure according to ISO 2921 and ASTM D1329.

The Elastocon TR Tester has 6 test stations, is computerized, and performs the test automatically after the cooling media has been cooled down, and the samples have been mounted. The instrument has an automatic release of the samples, after the pre-cooling time.

The computer controls both the temperature rise and measures the length change of the samples. The results are displayed in a graph and TR10, TR30, TR50 and TR70 values are calculated. The result can also be presented as a table with length change versus temperature. The TR-values and the table values can be exported to other software such as spreadsheets.

We recommend the automatic cooling system with liquid nitrogen that can be supplied.

Another useful option is the ET 04 freezer for precooling of the cooling media (usually ethanol), during nights or other times when the instrument is not in use. If the cooling media is pre-cooled, the amount of liquid nitrogen necessary to start the test will decrease remarkably.

Standards

The TR Tester performs well inside the apparatus requirements in ISO 2921 and ASTM D1329.

Temperature control

  • Temperature range: -80 °C to +20 °C
  • Accuracy: ± 0,5 °C
  • Resolution: ± 0,1 °C
  • Sensor: Pt 100, 1/3 DIN

Length measurement

  • Indicators: 6 pc digital encoders
  • Resolution: 0,1 mm
  • Measuring range: 0–150 mm

Important specifications

  • The rig is raised with the help of a pneumatic cylinder
  • Temperature controller with 0,1 °C setpoint and RS232 interface
  • Solid state relay for safe control
  • Temperature sensor close to the samples
  • Run-time meter (PLC)
  • The casing consists of steel, painted with epoxy paint

Included accessories with ET 01-II

  • Computer (Windows)
  • EC 12 software for the TR test
  • Support agreement 12 months
  • Manual in English
  • Necessary tools (e.g. stylus pen for PLC touch screen)
  • Accredited calibration including certificate

Options for ET 01-II

  • ET 01.02 – Automatic cooling with liquid nitrogen
  • ET 01.03 – Grips for O-rings (both whole and cut)
  • ET 01.04 – Extra Template 50 or 51 mm for setting the test distance

More information

Download product brochure: Elastocon Low Temperature Testers.pdf

Watch our video about the TR Tester: