Shear adhesion failure tester, SAFT, ES 07-II

The ES 07-II SAFT instrument is used to test hot-fail temperature in shear of hot melt adhesives.

Weights are attached to the shear samples and the temperature is increased with a controlled rate until the samples fail. The ES 07-II can test 6 samples simultaneously and have a timer for each test station that records the time to failure. This system has an improved sensor system for detecting the sample failure which works with small loads down to 100 g.

Standards

Complies to ASTM D4498 and other technical equivalent standards.

Temperature range

+30 °C to +200 °C

Temperature ramp, tolerance 0,5 °C

Important specifications

  • Inner chamber is made of stainless steel
  • Temperature controller with 0,1 °C set point
  • Solid state relay for safe control
  • Fixed over temperature fuse
  • Temperature controlled cooling fan for the electronics cabinet
  • Run-time meter

Included in purchase

  • ES 07-II instrument
  • 6 sets of weights, with 100 g, 500 g, and 1 kg weights included

More information

Product leaflet: Elastocon – Shear adhesion failure tester, SAFT, ES 07-II.pdf